beautypg.com

Calculate[1..n]:event – EXFO FTB-7000 OTDR Series for FTB-500 User Manual

Page 340

background image

SCPI Command Reference

326

OTDR

Product-Specific Commands—Description

The End of analysis event does not necessarily

correspond to the last event of a fiber link. It

indicates that the analysis has stopped before the

end of the link because the instrument has

reached the limit of its dynamic range.

In most cases, the OTDR analysis will return the

type of the last event as being either reflective or

non-reflective (event type 3 or 2).

Example(s)

CONF:ACQ:MODE ACQUISITION

INIT

INIT:STAT? Returns 0 when acquisition is

complete.

CALC:ANA TRC1

CALC:EVEN:COUN? TRC1 Ex.: Returns 4

(corresponding to 4 events).

CALC:EVEN? TRC1,1 (where 1 is the event

number. Values 1 to 4 are valid). Returns the

event corresponding to the specified number.

See Also

MMEMory[1..n]:LOAD:TRACe

TRACe[1..n]:CATalog?

:CALCulate[1..n]:EVENt?