Flintec FT-10xx User Manual
Page 76
![background image](/manuals/676078/76/background.png)
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014
Page 75 of 82
PLC Output to FT-10 CO Input RxPDO 1 (R_UL1)
Bit Number
RxPDO 1 (R_UL1) descriptions
D63 … D56
Set / Reset digital outputs
D55 … D48
Expanded Commands List ( Refer to Table 12.2)
D47 ... D43
Not in use
D42 … D38
Command
List
Bin
Dec
Commands
00000
0
None command is activated
00001
1
Zero
00010
2
Tare
00011
3
Clear
00100
4
00101
5
Adjust zero calibration
Calibration
00110
6
Adjust span calibration
( 1 )
00111
7
Total Load Cell Capacity
( 1 )
eCal
Coefficients
Refer to par.
[313]
01000
8
Average mV/V value
( 1 )
01001
9
Dead Load value
( 1 )
01010
10
Save the coefficients of eCal
01011
01110
11
14
Not used
01111
15
Set Point-1 Low
( 1 )
10000
16
Set Point-1 High
( 1 )
10001
17
Set Point-2 Low
( 1 )
10010
18
Set Point-2 High
( 1 )
10011
19
Set Point-3 Low
( 1 )
10100
20
Set Point-3 High
( 1 )
10101
21
Set Point-4 Low
(1)
10110
22
Set Point-4 High
(1)
10111
23
Set Point-5 Low
(1)
11000
24
Set Point-5 High
(1)
11111
31
Use the Expanded Command list ( Refer to Table 12.2)
D37 ... D33
Read Data
Selection
00000
0
Indicated weight
00001
1
Gross weight
00010
2
Tare weight
00011
3
Calibration Status ( Refer to Table 12.1)
00100
01110
4
14
Not used
01111
15
Set Point-1 Low
( 2 )
10000
16
Set Point-1 High
10001
17
Set Point-2 Low
( 2 )
10010
18
Set Point-2 High
10011
19
Set Point-3 Low
( 2 )
10100
20
Set Point-3 High
10101
21
Set Point-4 Low
( 2 )
10110
22
Set Point-4 High
10111
23
Set Point-5 Low
( 2 )
11000
24
Set Point-5 High
11111
31
Use the Expanded Command list ( Refer to Table 12.2)
D34
New CMD
Toggle
Apply commands which are listed in this table
D33~D0
D33~D37 bits defines the usage of this Dword.
(1) Write this
command with writing values to D0…D33 bits then apply New CMD.
(2) Only set point low addresses are used if the set point is programmed as standard.