FT-1x Quick Start, May 2012

FT-1x Quick Start

Important hint:
This instruction note does NOT replace the technical manual
It is only a brief survey for the commissioning of a type FT-1x
indicator.



1.

All mechanical assembly has to be done properly

2.

Execute electrical connections

Connect load cell with FT-1x:





Optional FT-1x to PC communication:

RS232,

PC typical allocation

Definition

D-Sub, 9-pole, female

Desk type housing and

panel type housing

Plug connector J11

Stainless steel housing

2

TxD

3

3

3

RxD

2

4

5

GND

7

2

Shield

Connector casing

1

3.

Setup the scale

1.

Turn on the instrument and wait for the start up.

2.

Access the menu for setup and calibration. Here you have
to press following keys in sequence:


and and ->0<- and and


With the <↑> and <↓> keys you select the parameter block [XXX], with the key you confirm
your selection. The key is used to exit the setup.

3.1

Weight display

Select parameter [212] and confirm with the key.

[CAP]

The capacity of the scale will be entered by the and keys.
The value will be confirmed by pressing the key.

[d]

The display resolution of the scale will be selected by the key.
The selection will be confirmed by pressing the key.

6-wire load cell

4-wire load cell

D-Sub, 9-pole, female

Desk type housing and

panel type housing

Plug connector J12

Stainless steel housing

+ Excitation

+ Excitation

1

1

+ Sense

+ Excitation

2

2

Shield

Shield

3

4

- Sense

– Excitation

4

6

– Excitation

– Excitation

5

7

+ Signal

+ Signal

7

3

– Signal

– Signal

8

5

Shield

Shield

Connector casing

4