2 fast access to the calibration, 3 exiting the programming and calibration – Flintec FT-10xx User Manual
Page 23
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014
5.2 Fast Access to the Calibration
The instrument has fast access calibration feature to earn time to the service technician. If only the
calibration adjustment is needed, follow the steps below to access the calibration parameters fast.
Display
Operation
[123.456 kg ]
Press
key until [ PASSWr ] prompts seen.
[PASSWr]
Press
+
+
keys sequentially.
[ --- ]
Press
key for confirm.
[ 310 ]
Zero Adjustment parameter.
“Calibration”
Press
key to start zero adjustment.
Or press
key to access span calibration without zero adjustment.
5.3 Exiting the Programming and Calibration
If you press
key on which parameter you are, you will get out of the active sub-block and reach the
next sub-block. If you press
key again, you will get out of the active block and reach the next main
block. If you press
key once again, the [ SAvE ] message appears on the display. Here you can press
key to save the changes into the memory, or you can press
key to store the changes until the
power goes off, or you can press
key to abort changes. [
Waıt ] message will be seen on the display
for a little while, and automatically get back to weighing mode.
Especially for legal metrological usage, please don’t forget to turn the power off and “OFF” position the
calibration DIP switch to start operation.