] filter, 3 parallel inputs and outputs [13, Page 36 – Flintec FT-10xx User Manual
Page 30
![background image](/manuals/676078/30/background.png)
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014
[12-] Filter
In this block the proper filter values according to the operating conditions can be entered. One of the
most important features of FT-1x series is viewing filter characteristic on the display and with the
help of this option, you can select the most suitable filter without exiting the programming mode.
[120 7 ] Filter
The filter value can be selected from 0 to 9 ( The minimum value of the filter at least 7 in normal
weighing applications). As you enter this parameter and press < Enter > key while [120 X] seen on
the display, the weight variation can be seen on the display. The value of the filter can be changed
by using <
> key and the weight variation for every value can be seen on the display. After
finishing the filter selection you can go to next step by pressing the < F > key.
5.4.3 Parallel Inputs and Outputs [13- ]
FT-10 has an optional up to 4 parallel inputs and up to 5 parallel outputs.
[130 0 ] Outputs
The outputs operate with respect to the weight value seen on the display.
At FT-10, FT-10 MB, FT-10 PB, FT-10 PN, FT-10 EN and FT-10 CO the outputs are programmed
as;
0 : No.
1 : ( Output1 = Sp1 ), (Output2 = Sp2 ), (Output3 = Sp3 ), (Output4 = Sp4 ), (Output5 = Sp5 )
2 : ( Output1 = Sp1 ), (Output2 = Sp2 ), (Output3 = Sp3 ), (Output4 = Sp4 ), (Output5 = Stable )
3 : ( Output1 = Sp1 ), (Output2 = Sp2 ), (Output3 = Sp3 ), (Output4 = Sp4 ), (Output5 = Error )
4 : Not used
5 : Functional outputs ( Refer to [ 7-- ] parameter group on
Page 36
At FT-10 AN the outputs are programmed as;
0 : No.
1 : ( Output1 = Sp1 ), (Output2 = Sp2 ), (Output3 = Sp3 ), (Output4 = Sp4 )
2 : ( Output1 = Sp1 ), (Output2 = Sp2 ), (Output3 = Sp3 ), (Output4 = Stable )
3 : ( Output1 = Sp1 ), (Output2 = Sp2 ), (Output3 = Sp3 ), (Output4 = Error )
4 : Not used
5 : Functional outputs ( Refer to [ 7-- ] parameter group on
Page 36
[131 0] Input 1
0 : Not used
1 : Zero
2 : Tare
3 : Clear 4 : Print
5 : Key lock
6 : Hold during active
7 : Peak during active
8 : Field bus input
[132 0] Input 2
0 : Not used
1 : Zero
2 : Tare
3 : Clear 4 : Print
5 : Key lock
6 : Hold during active
7 : Peak during active
8 : Field bus input
[133 0] Input 3 (only FT-10, FT-10 MB, FT-10 PB, FT-10 PN, FT-10 EN and FT-10 CO
0 : Not used
1 : Zero
2 : Tare
3 : Clear 4 : Print
5 : Key lock
6 : Hold during active
7 : Peak during active
8 : Field bus input
[134 0] Input 4 (only FT-10, FT-10 MB, FT-10 PB, FT-10 PN, FT-10 EN and FT-10 CO
0 : Not used
1 : Zero
2 : Tare
3 : Clear 4 : Print
5 : Key lock
6 : Hold during active
7 : Peak during active
8 : Field bus input