FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014
Page 71 of 82
12. CAN
OPEN
(
ONLY
FT-10
CO
)
After programming CANopen related parameters of the FT-10 PB indicator, you can communicate with the
instrument. EDS file is available on CD which is supplied together with the instrument.
Automatically detected and supported baud rates are 10 kbps, 50 kbps, 100 kbps, 125 kbps, 250 kbps, 500
kbps, 800 kbps, 1 Mbps, Autobaud (default).
[07-] CANopen
This sub-block includes the parameters related with the CANopen interfaces of FT-10 indicator.
[070 0] Data Format
0 : Signed 32 bit integer, no decimal point implied
1 : 32 bit float, decimal point implied
[071 000] Rack Address
The CANopen rack address of FT-10 will be entered via keypad between 001 to 126.
There are 2 LED’s near the CANopen connector which are;
Run LED
State
Indication
Comment
Off
Not online / No power
Check power and cable
Green
On-line, data exchange
-
Green, blinking
On-line, initializing
-
Green, single flash
Stopped
Check hardware damages
Green, flickering
Auto baudrate detection in progress
-
Red
CANopen configuration error
Check EDS file
Error LED
State
Indication
Comment
Off
-
No power or CANopen module is in
initialization state
Red, single flash
Warning limit reached
A bus error counter reached or
exceeded its warning level
Red, flickering
LSS
LSS services in progress
Red, double flash
Error count event
A guard- (NMT-Slave or NMT-
master) or heartbeat event
(Heartbeat consumer) has occurred.
Red
Bus off (Fatal Event)
Bus off.
1
Run LED
2
Error LED
3
CANopen interface