Flintec FT-10xx User Manual
Page 43
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014
Commands and Responses:
A
Read all weight data
Command : [ADR][A]
Response : [ADR][A][STATUS][SIGN][NET W][SIGN][TARE W][SIGN][GROSS W]
Example
:
Command : 01A
Response : 01AS+000123.4+000111.1+000234.5
01AD+000123.4+000111.1+000234.5
01AO
( ADC out error )
Comments :
The response is net, tare and gross weight values or error status.
All weight data is transmitted immediately after receiving the command.
B
Read Gross weight
Command : [ADR][B]
Response : [ADR][B][STATUS][SIGN][WEIGHT VALUE]
Example
:
Command : 01B
Response : 01BS+000123.4 (gross weight is stable and 123.4)
01BD+000123.4 (gross weight is dynamic and 123.4)
01B−
(under load)
Comments :
The response is the gross weight value (stable or dynamic) or error status.
Gross weight data is transmitted immediately after receiving command.
C
Clear the tare memory
Command : [ADR][C]
Response : [ADR][C][A] (Cleared and the scale is in gross mode)
Comments :
The response status is always Ack in weighing or force mode.
G
Read voltage value of DC power supply
Command : [ADR][G]
Response : [ADR][G][STATUS][VOLTAGE VALUE]
Example
:
Command : 01G
Response : 01GA234 (Power supply is 23.4 VDC)
01GA150 (Power supply is 15.0 VDC)
01GA090 (Power supply is 9.0 VDC)
Comments :
Voltage value is 3 byte and sends with 0.1 V increment.
I
Read indicated weight
Command : [ADR][I]
Response : [ADR][I][STATUS][SIGN][WEIGHT VALUE]
Example
:
Command : 01I
Response : 01IS+000123.4
(weight is stable and 123.4)
01ID+000123.4
(weight is dynamic and 123.4)
01I+
(overload)
Comments :
Indicated weight value (stable or dynamic) is transmitted immediately.
The weight value may be in gross or net.