1 ethernet parameters – Flintec FT-10xx User Manual
Page 67
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014
[036 0] Response Speed
0 : Modbus TCP Answer is sent immediately after Request is received.
1 : Modbus TCP Answer is delayed 20 msec after Request is received.
2 : Modbus TCP Answer is delayed 50 msec after Request is received.
This property is very helpful for slow PLC systems
11.1 Ethernet Parameters
There are 11 parameters for Ethernet network and Ethernet set up done by EtherX PC software over Local
Network Area as described in this section or You can entry from par. [ 03- ] blocks. EtherX PC software is
available in CD which is supplied together with the instrument.
Host Name
Device name of the instrument.
Default is ‘ ’.
IP Address
Obtain IP address manually.
Default is ‘192.168.16.250’.
Local Port
Ethernet connection port of the instrument.
Default is ‘502’.
Gateway
Network point that acts as an entrance to another network.
Default is ‘192.168.16.253’.
Subnet Mask
Describes IP address can be used in network.
Default is ‘255.255.255.0’.
Primary DNS
Obtain primary DNS manually.
Default is ‘208.67.222.222’.
Secondary DNS
Obtain secondary DNS manually.
Default is ‘208.67.220.220’.
Remote Connection
Automatic connection to any device on the network.
Default is ‘Disabled’.
Remote IP : IP address of the PC or Device to be connected automatically.
Remote Port : Ethernet connection point of PC or Device to be connected automatically.