Flintec FT-10xx User Manual
Page 50
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014
0
None
1
Zero
2
Tare
3
Clear
4
40010
R/W
20
Not used
40030
R/W
1
Calibratio
n
Dec
Description
0
None
188
Adjust Zero Calibration
220
Adjust Span Calibration
( 1 )
236
Total Load Cell Capacity
( 1 )
eCal
Coefficients
250
Average mV/V
( 1 )
171
Dead Load
( 1 )
23205
Save the coefficients of eCal
40031
R/W
2
Span Calibration Value / LC capacity / mV value / Dead load value
40033
R
1
Calibratio
n Status
D0 .. D7
Calibratio
n Process
Status
De
c
Description
1
Ready for calibration
3
Zero calibration in process …
4
Span calibration in process …
9
Error (Refer to D8 ... D15 )
D8 .. D15
Calibratio
n Errors
1
Calibration Timeout
- Restart calibration
2
ADC Error
- Re-energize the instrument
- If seen again, change the board.
3
Instrument can not be calibrating
- Check load cell cable
- Re-energize the instrument
34
Instrument can not be calibrating
- Load cell signal is very low or too high
35
Calibration Error
- Calibration loading is not enough
- Check test weight loading
- Check load cell connections
36
Calibration load value entry Error
- Test weight is too small. Increase the test
weight
37
Scale unstable
- Wait until scale become stable
- Check grounding wiring
38
The Calibration DIP switch is not ON
position.
- Check the calibration DIP switch.
40034
R/W
37
Not used
40071
R
1
Indicated weight
40072
R
1
Status
40073
R
1
Tare weight
40074
R
1
Gross weight
40100
R
1
Voltage of
Power Supply
Voltage of power supply is indicated with 0.1 V
increment.
For example: 23.4 VDC is indicated as integer 234
value.
40200
R
1
Status of Inputs
D0
Input-1