Table 10.2) – Flintec FT-10xx User Manual
Page 65
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014
Expanded Command List (always 32 bit integer)
The “D23 … D16” bits in 2
nd
Dword describes below.
Bit No
Description
D23…D16
D23…D16
Expanded
Commands
List
(continuation
of previous
page)
Bin
Dec
Commands
00000000
0
R
Voltage of
Power Supply
Indicated with 0.1 VDC
increment.
00000001
00111111
1
63
Not in use
01000000
64
RW
Filter
(1)
Refer to par. [120] on page 29
01000001
65
RW
Power On Zero
(1)
Refer to par. [202] on page 30
01000010
66
RW
Zeroing Range
(1)
Refer to par. [203] on page 30
01000011
67
RW
Auto Zero Tracking
(1)
Refer to par. [204] on page 30
01000100
68
RW
Tare
(1)
Refer to par. [205] on page 30
01000101
69
RW
Motion Detector
(1)
Refer to par. [206] on page 30
01000110
70
RW
Stability Period
(1)
Refer to par. [207] on page 30
01000111
71
RW
Operation mode
(1)
Refer to par. [210] on page 31
01001000
72
RW
Capacity
(1)
Refer to par. [212] on page 31
01001001
73
RW
Decimal point
(1)
0
XXXXOO
1
XXXXXO
2
XXXXXX
3
XXXXX.X
4
XXXX.XX
5
XXX.XXX
01001010
74
RW
Increment
(1)
1
X1
2
X2
3
X5
Table 10.2 - Expanded Command List
(1) Write this command after writing values to 1
st
Dword then apply this command with New CMD.
Programming steps of frequent used ……
Reading a weight value:
1. Compare
the D12…D15 bits of ‘FT-10 PN Output to PLC Input 2
nd
Dword’.
2. If there is not any error, read a weight value ( gross, net or tare),
Zero Calibration procedure:
1. Check the low byte of Calibration Status. it should be decimal '1' to start adjustment.
2. Write ‘Adjust Zero Calibration’ command and apply New CMD to start Zero calibration.
3. Check the low byte of Calibration Status. it is decimal '3' during zero calibration process.
4. The low byte of Calibration Status changes to decimal '1' at the end of the Zero calibration.
5. If the low byte of Calibration Status is '9', check the high byte of Calibration Status to understand the
calibration error.
Span Calibration procedure:
Check the low byte of Calibration Status. it should be decimal '1' to start adjustment.
Write ‘Adjust Span Calibration’ command after writing test weight values to 1
st
Dword, then apply this
command with New CMD to start Span calibration.
1. Check the low byte of Calibration Status. it is decimal '4' during span calibration process.
2. The low byte of Calibration Status changes to decimal '1' at the end of the Span calibration.
3. If the low byte of Calibration Status is '9', check the high byte of Calibration Status to understand the
calibre ation error.