2 setup software operating procedure, 1 operating procedure – KEYENCE N-400 User Manual
Page 42

Chapter 3 N-400 Setup Procedure
34
3
3.2
Setup Software Operating Procedure
3.2.1 Operating procedure
To set up the N-400 using the setup software, follow the procedure below.
1. Connect the N-400 to the host computer using the RS-232C cable.
➮
For the recommended cable, see 2.2.2 Connecting RS-232C.
2. Start the N-400 setup software.
3. Select a setup item, and set it to the desired condition.
4. Before sending the updated settings to the N-400, click on [COM PORT] to
match the communication parameters of the personal computer with the current
settings of the N-400.
The default settings of the N-400 are as follows:
• Baud rate: 9600 bit/s
• Data length: 7 bit
• Parity: Even
• Stop bit length: 1 bit
Immediately after you purchase the N-400, set [COM PORT] to the above
settings.
If you do not know the current settings of the N-400, press the N-400 test switch
for approximately 5 seconds. When the N-400 displays “S0(
50
)”, the N-400 is
automatically set to the above settings. The N-400 can communicate with the
host computer if the “COM PORT” settings are same as the above.
5. Click on [Transfer] to send the updated settings to the N-400.
(The current settings of the N-400 can also be read.)
6. Select [FILES] to save or print the settings, as required.
7. After the setup procedure is completed, click on [Exit] to close the setup soft-
ware.