82 Chapter
5
•
•
•
•
•
•
Restore Command
This command replaces the existing HI 6500 series parameters by the parameters stored on
a USB memory stick.
On the Web page, go the communication menu and left click on the Restore button.
On the HI 6500 series instrument display, select the Communication menu and use the
UP or DOWN button to select the USB submenu, then select the Restore menu item then
press the ENTER button.
Save Command
This command saves the current HI 6500 parameters to a USB memory stick.
On the Web page, go the communication menu and left click on the Save button.
On the HI 6500 series instrument display, select the Communication menu and use
the UP or DOWN button to select the USB submenu, then select the Save menu item
then press the ENTER button.
4-20 mA Communications Commands and Parameters
Available on the HI6510 only
The analog 4-20mA output is directly proportional to the programmed gross, net, or count
range; by setting the low value equal to 4mA and the high value equal to 20mA. These
values can be set from either the front panel or the web interface.
Weight Source Parameter
The weight source allows the 4-20mA output to be proportional to the gross weight, the net
weight, a count, or can be set to a known mA value
On the Web page, select the communication menu and using the weight source down list
select the required source for the 4 to 20mA output.