beautypg.com

Hardy HI 6500 Series User Manual

Page 78

background image

68 Chapter

5

Do Cal Lo Command

Before executing this Do Cal Lo command verify that the Cal Lo Weight value is

correct.

After setting/verifying the Cal Lo Weight value executing this command will run the

Hard Calibration process using the Cal Lo Weight, Cal Tolerance and Sensitivity
parameter settings.

On the Web page, left click on the Do Cal Lo button

On the HI 6500 series instrument display, select the calibration menu and use the

UP or DOWN button to select the Hard Cal submenu, then select the Do Cal Lo menu
item then press the ENTER button.

Do Cal Hi Command

Before executing this Do Cal Hi command verify that the Span Weight value is
correct. After setting/verifying the Span Weight value executing this command will
run the Hard Calibration process using the Cal Tolerance, Span Weight, and
Sensitivity parameter settings.

On the Web page, left click on the Do Cal High button as shown.

On the HI 6500 series instrument display, select the calibration menu and use the
UP or DOWN button to select the Hard Cal submenu, then select the Do Cal Hi menu
item then press the ENTER button.