Hardy HI 6500 Series User Manual
Page 88
78 Chapter
5
•
•
•
•
•
•
On the HI 6500 series instrument display, select the communication menu and use
the UP or DOWN button to select the Modbus-RTU submenu, then select the Baud
Rate menu item. The current Baud Rate value will be displayed, press the ENTER or
the RIGHT button if the value needs to be modified.
Parity Parameter
The Parity parameter has three options, EVEN, ODD, or OFF.
On the Web page, select the communication menu and then select the Modbus RTU
submenu, and left click inside the Parity field and select even, odd or off using the
drop down menu.
On the HI 6500 series instrument display, select the communication menu and use
the UP or DOWN button to select the Modbus-RTU submenu, then select the Slave
Address menu item. The current Parity value will be displayed, press the ENTER or
the RIGHT button if the Parity needs to be modified.