Zero command, Auto mode, Count operations parameters and commands – Hardy HI 6500 Series User Manual
Page 105
![background image](/manuals/657643/105/background.png)
Instrument Configuration
95
•
•
•
•
•
•
On the Web page, the value will appear on the Operations web page.
On the HI 6500 series instrument display, select the Operations menu and use the UP or
DOWN button to select the Zero Operations submenu, then select the Zero Amount menu
item. The current value will be displayed.
Zero Command
When the Zero command is selected the Gross weight value is set equal to 0 (zero) if the
Gross weight is within the Zero tolerance, and the displayed weight goes to 0.00 if the
Gross weight is being displayed.
On the Web page, select the Operations menu and left click on the Zero button
On the HI 6500 series instrument display, select the Operations menu and use the UP or
DOWN button to select the Zero Operations submenu, then select the Zero menu item and
press the ENTER button to execute this command.
NOTE
The Zero Command will fail if the scale is in motion
Auto Mode
Auto Mode enables one-button functionality when performing Zero and Tare operations,
automatically switching to Net mode when Tare is pressed. The default is Disabled.
On the Web page, select the Operations menu and left click to pull down the
Enabled/Disabled command to select Auto mode. Next select display mode Gross or Net.
Then click on the Save Parameters button.
On the HI 6500 series instrument display, select the Operations menu and use the UP or
DOWN button to Enable or Disable Automode. For display mode, select Gross or Net.
Press the ENTER button to execute this command.
Count Operations Parameters and Commands
The count mode allows the unit to be calibrated and used as a counter. This mode allows
the user to "count" the number of pieces placed on the scale.