104

720i Installation Manual

See descriptions below

Bit fields. Comma-separated sequence of bit field specifiers. Must
be exactly 8 bits. Minus sign ([–]) inverts the bit.

B0

Always 0

B1

Always 1

B2

Configuration

=1 if even parity

B3

Dynamic

=1 if MODE=NET

B4

Dynamic

=1 if COZ

B5

Dynamic

=1 if standstill

B6

Dynamic

=1 if gross negative

B7

Dynamic

=1 if out of range

B8

Dynamic

=1 if secondary/tertiary

B9

Dynamic

=1 if tare in system

B10

Dynamic

=1 if tare is keyed

B11

Dynamic

=00 if MODE=GROSS
=01 if MODE=NET
=10 if MODE=TARE
=11 (not used)

B12

Dynamic

=00 if UNITS=PRIMARY
=01 if UNITS=SECONDARY
=10 if UNITS=TERTIARY
=11 (not used)

B13

Configuration

=00 (not used)
=01 if current DSPDIV=1
=10 if current DSPDIV=2
=11 if current DSPDIV=5

B14

Configuration

=00 (not used)
=01 if primary DSPDIV=1
=10 if primary DSPDIV=2
=11 if primary DSPDIV=5

B15

Configuration

=00 (not used)
=01 if secondary DSPDIV=1
=10 if secondary DSPDIV=2
=11 if secondary DSPDIV=5

B16

Configuration

=00 (not used)
=01 if tertiary DSPDIV=1
=10 if tertiary DSPDIV=2
=11 if tertiary DSPDIV=5

B17

Configuration

=000 if current DECPNT=8888800
=001 if current DECPNT=8888880
=010 if current DECPNT=8888888
=011 if current DECPNT=888888.8
=100 if current DECPNT=88888.88
=101 if current DECPNT=8888.888
=110 if current DECPNT=888.8888
=111 if current DECPNT=88.88888

Format Identifier

Defined By

Description

Table 11-8. Custom Stream Format Identifiers