Wake from storage routine – Bio-Rad Bio-Plex Manager™ MP Software Upgrade User Manual
Page 88

7
| Maintenance Routines
86
| Bio-Plex Manager MP Software
Wake from Storage Routine
Description of Routine
The Wake from Storage routine prepares the instrument to be used after it has been
idle for a long period of time.
When to Run this Routine
Run the Wake from Storage routine if the instrument has been idle for 2 weeks or
longer or if it has been prepared for storage.
Bio-Plex
Manager
MP adds Wake From Storage to the Daily Start Up routine if it
detects that the instrument has been idle for 2 weeks or longer. Therefore, if you run
the Daily Start Up routine daily, you will be running this routine as recommended.
Estimated Run Time
2 minutes
To Run the Wake From Storage Routine
1.
Navigate to the Maintenance view.
2.
In the Select Routine section, select Wake From Storage from the Additional
Routines dropdown list.
3.
Add 5 ml distilled or deionized water and 5 ml 70% isopropanol to the
appropriate reservoirs in the reagent block.
4.
Click Start.