Unclog routine – Bio-Rad Bio-Plex Manager™ MP Software Upgrade User Manual

Page 86

background image

7

| Maintenance Routines

84

| Bio-Plex Manager MP Software

Unclog Routine

Description of Routine

This procedure performs a series of fluidics operations to remove possible

obstructions from the fluidics lines.

When to Run this Routine

If you suspect there is an obstruction in the MAGPIX system, run the Unclog routine.

Estimated Run Time

2 minutes

To Run the Unclog Routine

1.

Navigate to the Maintenance view.

2.

In the Select Routine section, select Unclog from the Additional Routines
dropdown list.

3.

Add 5 ml distilled or deionized water and 5 ml 0.1N NaOH to the appropriate
reservoirs in the reagent block.

4.

Insert the plate in the microplate platform and click Start.