Well Types

The following are different well types that you can specify on the plate:

Blank

Standard Undefined

Control Unknown

Blank Wells

In certain types of assays, such as the Bio-Plex phosphoprotein assay, it might

be useful to subtract the assay background from the readings of standards,

controls, and unknown samples. Prepare blank wells that contain all of the

assay components excluding the sample. Blank wells are read along with the

rest of the assay. Bio-Plex

Manager 6.x then subtracts the mean background

reading of these wells from the fluorescence intensity values of the wells

containing standards, controls, and unknowns.

Control Wells

Control wells contain samples of known concentration. The observed

concentration of the control wells is compared to the expected concentration

and calculated at the end of the reading.

Standard Wells

Standard wells contain analytes of known concentration. A series of known

concentrations of an analyte is used to generate a standard curve of

fluorescence intensity compared to analyte concentration. The regression

equation for the curve is used to calculate the concentration of analytes in the

unknown samples and controls.