Daily Maintenance Routines

User Guide

| 59

Daily Maintenance Routines

Bio-Rad recommends that you run the following routines each day you use the

MAGPIX instrument.

At the Beginning of the Day

Run the routine in the Recommended Routine section. One or both of the following

routines are recommended each day:

Daily Start Up

Enhanced Maintenance

Continue to run the recommended routine until maintenance is up to date and None

is displayed in the Recommended Routine section.

At the End of the Day

At the end of each day, run the Shut Down routine before you shut down the

instrument.

Daily Start Up Routine

The Daily Start Up routine runs different procedures depending on the state of the

instrument. Bio-Plex

Manager

MP detects whether the instrument needs to be

verified, calibrated, or both, and it runs the appropriate commands as part of Daily

Start Up. This routine can take 4–12 minutes.

Note:

When you turn on the MAGPIX instrument, Daily Start Up is generally the

routine that is recommended. However, there may be times when Enhanced

Maintenance is recommended, for example, if calibration and verification

previously failed. Run whatever routines are recommended until maintenance is

up to date and no routines appear in the Recommended Routine section.

Tip:

If verification, calibration, or both routines fail after successive attempts,

contact Bio-Rad Technical Support for assistance.