Enhanced maintenance, Shut down routine, Running a routine – Bio-Rad Bio-Plex Manager™ MP Software Upgrade User Manual

Page 62: Enhanced maintenance shut down routine

background image

6

| Maintaining the MAGPIX System

60

| Bio-Plex Manager MP Software

Enhanced Maintenance

The Enhanced Maintenance routine will be recommended, for example, if calibration

or verification failed on the previous day. Once this routine is successfully

completed, it is followed with a recommendation to run the Daily Start Up routine.

Shut Down Routine

If you do not run the Shut Down routine at the end of the previous day, when you

start up the instrument the next day, the Status section in the Maintenance view

shows that Shut Down is not current. Run the Shut Down routine first. This routine

takes 2 minutes. Once that routine has successfully completed, then run the routine

in the Recommended Routine section.

Running a Routine

In addition to running the recommended routines as part of daily maintenance, all

routines may be selected and run at any time.

To run a routine

1.

From the Maintenance view, select the routine from the Select Routine list.

2.

If the routine does not appear in the list, click the Additional Routines dropdown
list and select the desired routine.

3.

Fill the reservoirs as indicated in the Reagent Block.

The Reagent Block is colored with the reservoirs and wells that need to be filled

for the selected routine. For more information on filling the reservoirs, see

Setting Up the Reagent Block on page

64

.

4.

If indicated in the Reagent Block, fill the well strip with the appropriate
reagents.

For more information on filling the well strip, see

Setting Up the Reagent Block

on page

64

.

5.

If the routine involves calibration and/or verification, select the kit in the Select
Kit section.