Verification routine – Bio-Rad Bio-Plex Manager™ MP Software Upgrade User Manual
Page 87

Verification Routine
User Guide
| 85
Verification Routine
Description of Routine
The Verification routine ensures that the calibration settings are valid and ensures
that well-to-well cross talk does not occur.
When to Run this Routine
Run the Verification routine when its status in the Maintenance view is not up to
date. Bio-Plex
Manager
MP adds verification to the Daily Start Up routine in
accordance with the manufacturer’s recommendation. Therefore, if you run the Daily
Start Up routine daily, you will be running the Verification routine as recommended.
Estimated Run Time
4 minutes
To Run the Verification Routine
Before you run the Verification routine, you must import the verification kit
information. For more information, see
Calibration and Verification Kits on page
.
1.
Navigate to the Maintenance view.
2.
In the Select Routine section, select Verification from the Additional Routines
dropdown list.
3.
Add 5 ml 70% isopropanol to the appropriate reservoir in the reagent block.
4.
Vortex VER, Fluidics1, and Fluidics2 each for 10 sec.
5.
Add 6 drops each of CAL, VER, Fluidics1, and Fluidics2 to the indicated wells.
6.
Click Start.