Bio-Rad Bio-Plex Manager™ MP Software Upgrade User Manual

Page 78

background image

7

| Maintenance Routines

76

| Bio-Plex Manager MP Software

Estimated Run Time

10 minutes

To Run the Calibration and Verification Routine

Before you run the Calibration and Verification routine, you must import the

calibration and verification kit information. For more information, see

Calibration and

Verification Kits on page

66

.

1.

Navigate to the Maintenance view.

2.

In the Select Routine section, select Calibration and Verification from the
Additional Routines dropdown list.

3.

Add 5 ml 70% isopropanol to the appropriate reservoir in the reagent block.

4.

Vortex CAL, VER, Fluidics1, and Fluidics2 each for 10 sec.

5.

Add 6 drops each of CAL, VER, Fluidics1, and Fluidics2 to the indicated wells.

6.

Click Start.