Removing the Well Formatting
Removing the formatting of a well returns it to its unformatted state. Unformatted
wells are not read by the array reader.
To remove the formatting of a well
1.
Click Undefined.
2.
Click and drag your cursor over the wells you want to clear.
The formatting is removed and the wells return to their unformatted state.
Changing the Well Formatting
You can overwrite the formatting of a well by simply applying another sample type
over the wells.
To change the formatting of a well
1.
Click the button for the sample type you want to use.
2.
Click and drag your cursor over the wells whose format you want to change.
The wells are now formatted with the new well type.
About Standards
Standards are analytes of known concentration. Standards are used to generate a
standard curve of values using one of the several regression methods included in
Bio-Plex
Manager
6.x. This curve is used to calculate concentrations of your
unknowns.
In Bio-Plex
Manager
MP, you define your standard lot, including the starting
concentration for each of the analytes. This definition is stored and it can be reused
each time you run a plate with those analytes.