3

| Running Protocols

28

| Bio-Plex Manager MP Software

Using the Plate Heater

For some experiments, for example nucleic acid testing, you may need to heat and

maintain the temperature of the plate with the MAGPIX plate heater. You can turn

the heater on from any view in Bio-Plex Manager MP and set the temperature of the

plate from 35–60ºC in 0.5º increments.

Note:

While the plate temperature is being set, you cannot run your protocol.

To use the plate heater

1.

In the dashboard, select the “Plate heater off” checkbox to turn the heater on.

The “Turning on” message appears. When the heater is turned on, Set Temp is

displayed.

2.

Use the left- and right-pointing triangles to set the plate temperature.

The temperature can be set from 35–60ºC in 0.5º increments.

When the temperature of the plate falls within range of the set temperature, the

thermometer turns green.

3.

When you are done, select Set Temp to turn the heater off.

Running the Protocol

After you create your protocol, you are ready to run it. Refer to

Chapter 2, Creating

Protocols

for more information on creating protocols.

Note:

You can run only protocols created in Bio-Plex

Manager

MP. You cannot

run protocols created in other versions of Bio-Plex

Manager such as

Bio-Plex

Manager

6.1.

The data from each run are stored as a result in the Bio-Plex

Manager

MP database.

Result names must be unique. You can retrieve the data by exporting the result to

Bio-Plex

Manager

6.x.