6

| Maintaining the MAGPIX System

66

| Bio-Plex Manager MP Software

To set up the Reagent Block

1.

Fill any colored reservoir with the indicated liquid.

2.

For each colored well in the well strip, locate the appropriate calibration or
verification kit.

3.

Vortex each bottle for at least 10 sec.

4.

Fill the well with 6 drops of the solution from the bottle with the matching cap
color.

Calibration and Verification Kits

Calibration and verification kits are sold separately. Each calibration kit comes with

the following:

1 bottle of MAGPIX calibrator

1 calibration kit CD

1 pack of well strips

Each verification kit comes with the following:

1 bottle of MAGPIX verifier

1 bottle of MAGPIX fluidics 1

1 bottle of MAGPIX fluidics 2

1 verification kit CD

1 pack of well strips

About Kit Lot Numbers

The lot number on the box identifies the kit. Use this lot number to coordinate the

reagents with the calibration or verification values (on the CDs). The bottle for each

reagent is not identified with the lot number for the kit. Each bottle is identified with

its own lot number, but this number is different from the lot number for the kit.