Calibration and verification kits, Calibration and – Bio-Rad Bio-Plex Manager™ MP Software Upgrade User Manual
Page 68

6
| Maintaining the MAGPIX System
66
| Bio-Plex Manager MP Software
To set up the Reagent Block
1.
Fill any colored reservoir with the indicated liquid.
2.
For each colored well in the well strip, locate the appropriate calibration or
verification kit.
3.
Vortex each bottle for at least 10 sec.
4.
Fill the well with 6 drops of the solution from the bottle with the matching cap
color.
Calibration and Verification Kits
Calibration and verification kits are sold separately. Each calibration kit comes with
the following:
1 bottle of MAGPIX calibrator
1 calibration kit CD
1 pack of well strips
Each verification kit comes with the following:
1 bottle of MAGPIX verifier
1 bottle of MAGPIX fluidics 1
1 bottle of MAGPIX fluidics 2
1 verification kit CD
1 pack of well strips
About Kit Lot Numbers
The lot number on the box identifies the kit. Use this lot number to coordinate the
reagents with the calibration or verification values (on the CDs). The bottle for each
reagent is not identified with the lot number for the kit. Each bottle is identified with
its own lot number, but this number is different from the lot number for the kit.