Shut Down Routine

User Guide

| 83

Shut Down Routine

Description of Routine

The Shut Down routine cleans the fluidics lines and prevents a buildup of debris

within the system.

When to Run the Routine

Bio-Rad recommends that you run the Shut Down routine at the end of each day

that you use the instrument. Refer to the Shut Down entry in the status section in

the Maintenance view to determine whether the Shut Down has been run that day.

Estimated Run Time

2 minutes

To Run the Shut Down Routine

1.

Navigate to the Maintenance view.

2.

In the Select Routine section, select Shut Down.

3.

Add 5 ml distilled or deionized water and 5 ml bleach to the appropriate
reservoirs in the reagent block.

4.

Click Start.