3

| Running Protocols

30

| Bio-Plex Manager MP Software

If you select a pre-run routine, refer to the Reagent Block diagram and fill the

appropriate reservoirs with the specified fluids.

6.

(Optional) In the Post-Run Routine list, select any routines you want executed
after the protocol run.

Note:

Bio-Rad recommends you run the Wash Between Plates routine

after each protocol run to prevent traces of sample or other debris from

collecting inside the Bio-Plex® suspension array system.

If you select a post-run routine, refer to the Reagent Block diagram and fill the

appropriate reservoirs with the specified fluids.

7.

Accept the default for “Stop runs with low bead count” or clear the checkbox if
you do not want the run interrupted.

By default, the run is interrupted if three consecutive wells have a bead count of

less than 20.

8.

Click Create/Run Protocols.

The plate retracts and Bio-Plex

Manager

MP starts the run.