Clear Bubbles Routine

User Guide

| 77

Clear Bubbles Routine

Description of Routine

The Clear Bubbles routine performs a series of fluidics operations to force out air

bubbles that may be trapped in the cuvette.

When to Run this Routine

If you suspect there is air in the MAGPIX system, run the Clear Bubbles routine.

Estimated Run Time

3 minutes

To Run the Clear Bubbles Routine

1.

Navigate to the Maintenance view.

2.

In the Select Routine section, select Clear Bubbles from the Additional
Routines dropdown list.

3.

Add 5 ml distilled or deionized water and 5 ml 70% isopropanol to the
appropriate reservoirs in the reagent block.

4.

Click Start.