Running a Routine

User Guide

| 61

a.

Click the Calibration Kits dropdown arrow to select the correct calibration
kit.

b.

Click the Verification Kits dropdown arrow to select the correct verification
kit.

Note:

The lot number on the MAGPIX Calibration Kit or the MAGPIX

Verification Kit CD must match the lot number of the kit to which the

reagents belong.

For more information on calibration and verification kits, see

Calibration and

Verification Kits on page

66

.

6.

Click Start.

Once the routine is completed, a dialog box appears informing you that the

routine is completed.

Note:

You cannot run a protocol while a routine is running. You must wait for

the routine to complete.