Editing a Program
With the program edit function, you can edit the registered ramp time, ramp level, soak time, wait
function, repeat start segment and the number of repeat cycles on the segment basis.
CAUTION : You cannot delete the registered segments or the repeat command and insert new segments or
a repeat command.
Based on Figure 4 of “Inputting a Program,” this paragraph discusses how to edit Segment 3 from the
step operation to the constant ratio ramp operation.
Indication after Step/Step Procedure
Explanation
1. Load the program number that you want to
edit.
Push either the ▼ key or the ▲ key after pushing the
MODE key.
•
Push the MODE key to activate the function selection
mode, and push either the ▼ key or the ▲ key.
⇒
⇒
⇒
⇒
The controller turns to the program input selection
mode. The main display flashes
.
2. Push
the
ENTER key
•
Push the ENTER key.
⇒
⇒
⇒
⇒
The controller turns to the program input/edit mode.
The main display flashes
.
3. Push
the
▲
key
Ex., Editing Program 3
•
Push the ▲ key to flash the program number that you
want to edit on the main display (
in this
example).
⇒
⇒
⇒
⇒
On the sub display, the left two digits show the
number of used segments while the right two digits
show the number of unused segments.
4. Push
the
ENTER key
•
Push the ENTER key.
⇒
⇒
⇒
⇒
The main display flashes the first registered segment
number of Program 3.
⇒
⇒
⇒
⇒
On the sub display, the left digit shows the program
number that you are working on, and the right digit
shows the number of unused segments.
5. Push
the
▲
key
Load Segment 3
•
Push the ▲ key to flash the segment number that you
want to edit on the main display (
in this
example).
(Ex. 3 unused segments)
(Ex . 3 Programs registered)
(Ex. 2 used segments)
(Ex, 0 unused segments)
(Ex. 5 used segments)
(Ex. 0 unused segments)
(Ex., the program number that
you are working on)
(Ex, 0 unused segments)
(Ex., the program number that
you are working on)