Contents of a Segment

Other Elements relating to Segment

1. Wait Function (See the following example for setting the wait function).

If unexpected additional load or disturbance prevents measured temperatures from reaching the desired
temperature, this function keeps the operation in a block until the desired temperature is achieved. After
inputting the soak time, set ON or OFF.

2. End Segment

At the end of inputting a new program, set

in the ramp time of the last segment. When you have

composed a program up to the fifth segment, you do not need to set

Turning ON the wait function

3. Hold

If you set

in the soak time, the controller holds the desired temperature (ramp level) and closes the

program input mode.

If

is set, the holding time between 0 minute 0 second and 999 minutes flashes when you select the

soak time.

Indication when you have selected “Hold.”

! Be aware that this is different from the temperature holding function in the function menu.

Ramp Level

Push either the ▼ key
or ▲key.

999 minutes