The display after the power restoration

when the resume function is in activity.

If a blackout, which includes a instantaneous blackout, occurs during the quick auto-stop operation
and the program operation when the resume function has been set to on, after the power restored, the
sub display flashes the information of having a blackout after the quick auto-stop operation or during
the program operation. If you turn off and on the power switch (circuit breaker) during the quick
auto-stop operation or the program operation, the same screen appears.

Indication after Step/Step Procedure

Explanation

In case that a blackout occurs during the quick auto-stop
operation or the quick auto-stop operation and then
HC200 goes on running the operation after the power
restoration / In case that a blackout occurs during the
program operation and then HC200 goes on running the
fixed temperature operation at the preserved
temperature after the power restoration

The sub display flashes

whatever the screen

has appeared on it. During the quick auto-stop
operation, you can switch the screen on the sub
display from

to the preset temperature or the

remaining time before the operation stop by using the
sub display screen switching function.(

)

You can also select and set the other function modes

during the quick auto-stop operation or the program
operation.

In case that a blackout occurs during the quick auto-stop
operation and then the quick auto-stop operation has
come to an end after the power restoration

Ordinarily, the sub display flashes

when

finished the quick auto-stop operation. However, if a
blackout occurs during the operation, the sub display
flashes

(Abnormal end)

You can select and set the function modes other than

mode.