4 Configuration table
66
4.2.5
Configuration level 3 (C3-level)
Parameter
Value/selection
Description
Relay output
Configuration level 3
➔ Relay output
X Relay
1 — 8
Select relay
Status:
OFF, ON
OFF = relay inactive
ON = relay operates as an
LK output
Mathematical
module
Configuration level 3
➔ Mathematical module
Mathematical m.:
OFF, ON
Activate/de-activate
mathematics module
Interface
Configuration level 3
➔ Interface
Protocol:
J-BUS, MOD-BUS
Set up RS422/RS485
Select protocol
Baud:
0.15, 0.3, 1.2, 2.4, 4.8, 9.6,
19.2, 38.4, 125,
187.5kBaud
Select baud rate
Data format:
8/1/NO,
8/1/ODD,
8/1/EVEN,
8/2/NO,
8/1/ZERO
Select data format
(Data bits / stop bits /
parity)
Address:
1 — 255
Select address
Min. response time:
0 — 500msec.
Select minimum response
time
External text
Configuration level 3
➔ External text
X External text
1 — 8
Select the number of the
external text
Contact:
OFF, Log. input 1,
Log. input 2, Log. input. 3,
Log. input 4, Log. input 5,
Log. input 6, Log. input 7,
Log. input 8
Select logic input.
(if OFF: the text entry is not
required)
Text:
16 characters
Enter text
Binary-linked
external text
(text logic links)
Configuration level 3
➔ Text logic links
Link:
OFF, Log. input 1,
Log. input 1-2,
Log. input 1-3,
Log. input 1-4
Select the inputs which are
to be binary-linked, (if OFF,
all further sub-parameters
are not required)