3 update commands, Table 5-20 – INFICON Composer Elite Gas Concentration Monitor User Manual
Page 95
5 - 29
PN
07
4-
56
6-
P1
B
Composer Elite Operating Manual
5.6.3 Update Commands
Message = U
Command ID =
Sensor =
Internal Parameter =
.
Response = None (header and trailer only)
Table 5-20 Query & Update commands
Q & U
Command
ID
Name Range
(low/high)
Data
Type
Note
Restrictions
2
Carrier Molecular Weight
1.000 -
1000.000
(AMU)
Float
This value represents the
molecular weight of the lighter gas.
-
3
Carrier Gamma
1.0000 -
2.0000
Float
This value represents the
Carrier’s specific heat ratio.
-
4
Precursor Molecular Weight
1.000 -
1000.000
(AMU)
Float
This value represents the
molecular weight of the
heavier gas.
-
5
Precursor Gamma
1.0000 -
2.0000
Float
This value represents the
Precursor’s specific heat ratio.
-
6
Allow User Zero
0-1
Integer Current value of Allow User Zero.
-
7
Reserved For Future Use
8
Reserved For Future Use
9
Query Zone 1 Temperature
30.00 -
65.00(ºC)
Float
Current value of the Zone 1
Temperature Setpoint parameter of
the Acoustic Sensor’s resonant
chamber.
-
Zone 2 Delta
-99.999 -
99.999(ºC)
Float
Current value of the Zone 2
Temperature Delta parameter of
the Acoustic Sensor’s resonant
chamber.
Will reject an
update for
Heater 2 if the
temperature is
not within the
30 to 68°C
range.
10
Result Averaging Filter Depth
0-100
Integer Current value of Averaging Filter
Depth.
-
20
Sets/Gets Bit Pattern for S0
Command
0 - 0x7FF8
0000
Encode Each bit represents a piece of data
returned. When a bit is set (1) that
piece of data is returned in the S0
command response. Refer to
for details of which byte
relates to what data.
Bit = 0 Data not sent.
Bit = 1 Data is sent.