Concentration drift / offset

Reference Zero was
performed before Acoustic
Sensor was completely
purged with a pure carrier
gas (refer to

section

4.1.10, What Happens if
the Reference Zero is
Improperly Set?, on page
4-8

).

Concentration offset is
equal to the residual
precursor concentration at
the time the Reference
Zero was performed.

Thoroughly purge
Acoustic Sensor with pure
carrier gas and then
perform Reference Zero
(refer to

section 3.2,

Calibration, on page 3-2

).

Large barometric pressure
change without performing
Reference Zero.

Depends on magnitude of
barometric pressure
change.

Perform Reference Zero
(refer to

section 3.2,

Calibration, on page 3-2

).

Reference Zero performed
at inappropriate time.

Depends on magnitude of
temperature change after
Reference Zero was
performed.

Wait for STEADY and AT
TEMP indicators to appear
and then allow additional
time for maximum
temperature stability
before setting Reference
Zero.

Change of process
pressure.

Refer to

section 4.1.6,

What is the Effect of
Pressure Variation?, on
page 4-4

.

Perform Reference Zero
(refer to

section 3.2,

Calibration, on page 3-2

).

Change of process flow.

Refer to

section 4.1.7,

What is the Effect of Flow
Variation?, on page 4-5

.

Perform Reference Zero
(refer to

section 3.2,

Calibration, on page 3-2

).

Table 8-1 FMEA (continued)

Symptom

Possible Cause

Effect on Concentration
Accuracy

Remedy