INFICON Composer Elite Gas Concentration Monitor User Manual
Page 82
5 - 16
PN
07
4-
56
6-
P1
B
Composer Elite Operating Manual
Sensor 3
SS, LS, MH1/PC,
MH2/UZ, AT, CS,
ER, RS, BM, WA
Refer to
Integer
Status bits Refer to
.
SS bit is not relevant in the data set format.
Sensor 3
Measurement
Mode
IDLE = 0
READY =1
SEARCH = 2
TRACK = 3
QTRACK = 4
BASELINE = 6
Encoded
Current System Measurement Mode.
Sensor 3
Concentration
0.0 – 100.0
mole %
Float
The current concentration value.
Sensor 3
Temperature 1
30.0 – 65.0 C
Float
The current temperature value of first
temperature sensor.
Sensor 3
Temperature 2
30.0 – 65.0 C
Float
The current temperature value of second
temperature sensor.
Sensor 3
Heater 1 Status
0 - 6
Short
0 = Off
1 = Not at temp
2 = At temp
3 = Failed
4 = Cable Disconnected
5 = Near Temp
6 = Over Temp
Sensor 3
Heater 2 Status
0 - 6
Short
0 = Off
1 = Not at temp
2 = At temp
3 = Failed
4 = Cable Disconnected
5 = Near Temp
6 = Over Temp
Sensor 3
Sample Number
0 – 255
Byte
The current sample number.
Sensor 4
SS, LS, MH1/PC,
MH2/UZ, AT, CS,
ER, RS, BM, WA
Refer to
Integer
Status bits Refer to
.
SS bit is not relevant in the data set format.
Sensor 4
Measurement
Mode
IDLE = 0
READY =1
SEARCH = 2
TRACK = 3
QTRACK = 4
BASELINE = 6
Encoded
Current System Measurement Mode.
Sensor 4
Concentration
0.0 – 100.0
mole %
Float
The current concentration value.
Sensor 4
Temperature 1
30.0 – 65.0 C
Float
The current temperature value of first
temperature sensor.
Sensor 4
Temperature 2
30.0 – 65.0 C
Float
The current temperature value of second
temperature sensor.
Table 5-14 All sensor card data set (continued)
Name
Range
Data Type Note