beautypg.com

2 composer elite installation, As described by – INFICON Composer Elite Gas Concentration Monitor User Manual

Page 36

background image

2 - 8

PN

07

4-

56

6-

P1

B

Composer Elite Operating Manual

2.2 Composer Elite Installation

2.2.1 Installing The Sensor Control Unit in a 19 in. (48.26 cm) Rack

Install the Sensor Control Unit in a standard 48.26 (19 in.) rack using the optional
Full Rack Extender (PN 782-900-007).

1

Assemble the two 7.62 x 12.7 cm (3 x 5 in.) Extender Kit side panels and the
larger front and rear panels into a box configuration using the eight 6-32
flat-head screws.

2

Thread two 10-32 shoulder screws from the inside of one of the box sides until
the threads extend fully to the outside.

3

Attach the Extender Kit to the Sensor Control Unit by threading the shoulder
screws into the matching holes in the Sensor Control Unit covers.

4

Attach the rack mounting ears with the 10-32 flat-head screws provided.

5

Carefully lift the assembly into a full width, 13.33 cm (5-1/4 in.) high rack space.

6

Attach the assembly to the rack using the necessary screws (not provided).

7

Connect the Sensor Control Unit to earth ground as described by

section

2.1.2.3, Composer Elite System Grounding Requirement, on page 2-4

.

CAUTION

Connect the Sensor Control Unit to earth ground as
described by

section 2.1.2.3, Composer Elite System

Grounding Requirement, on page 2-4

.