2 composer elite installation, As described by – INFICON Composer Elite Gas Concentration Monitor User Manual
Page 36
2 - 8
PN
07
4-
56
6-
P1
B
Composer Elite Operating Manual
2.2 Composer Elite Installation
2.2.1 Installing The Sensor Control Unit in a 19 in. (48.26 cm) Rack
Install the Sensor Control Unit in a standard 48.26 (19 in.) rack using the optional
Full Rack Extender (PN 782-900-007).
1
Assemble the two 7.62 x 12.7 cm (3 x 5 in.) Extender Kit side panels and the
larger front and rear panels into a box configuration using the eight 6-32
flat-head screws.
2
Thread two 10-32 shoulder screws from the inside of one of the box sides until
the threads extend fully to the outside.
3
Attach the Extender Kit to the Sensor Control Unit by threading the shoulder
screws into the matching holes in the Sensor Control Unit covers.
4
Attach the rack mounting ears with the 10-32 flat-head screws provided.
5
Carefully lift the assembly into a full width, 13.33 cm (5-1/4 in.) high rack space.
6
Attach the assembly to the rack using the necessary screws (not provided).
7
Connect the Sensor Control Unit to earth ground as described by
2.1.2.3, Composer Elite System Grounding Requirement, on page 2-4
CAUTION
Connect the Sensor Control Unit to earth ground as
described by
section 2.1.2.3, Composer Elite System