1.5.10 Perimeter for Maintenance Access

Acoustic Sensor to be removed from the process gas tubing (Swagelok VCR-4
fittings) require side-to-side or top-to-bottom clearance of 16 cm (6 in.)
for wrench rotation.

Interconnect Cable to be removed or replaced from either Acoustic Sensor or
Sensor Control Unit requires 8 cm (3 in.) clearance.

Sensor Control Unit to be removed from rack, full 30.5 cm (12 in.)
length required.

1.5.11 Environmental Requirements

The following detail Composer Elite environmental requirements.

Operation . . . . . . . . . . . . . . . . . . . . . Composer Elite is designed for

indoor use only.

Altitude Range . . . . . . . . . . . . . . . . . Composer Elite can be used to a maximum

altitude of 2000 m (6561 ft.).
For applications above this altitude, please
contact INFICON.

Maximum Humidity. . . . . . . . . . . . . . Non-condensing relative humidity at

operating temperature.

Pollution Degree. . . . . . . . . . . . . . . . Category II (as defined by EN61010-1, only

non-conductive pollution occurs).

Operating Temperature . . . . . . . . . . Refer to

section 1.5,