1.4.2 Sensor Control Unit Power-Up Verification

It is not necessary to have Acoustic Sensors or communications cables connected
to the Sensor Control Unit at this time.

1

Set the power switch on the rear panel of the Sensor Control Unit to the 0 (OFF)
position. (See

Figure 3-8 on page 3-14

.)

2

Using the INFICON (optional) 24 V power supply, or a power supply as defined
in

section 1.5.4

and

Table 1-1 on page 1-12

, insert the output cable of the 24 V

power supply into the connector labeled 150W 24V on the rear panel of the
Sensor Control Unit (see

Figure 3-8 on page 3-14

). Make certain the cable is

inserted until it locks in place.

3

Connect mains power (refer to

section 1.5.4.2 on page 1-12

)

to the input of the 24 V power supply.

4

Power up the Sensor Control Unit from the rear panel by setting the power
switch to the 1 (ON) position. The Boot screen will appear followed by a
System Overview screen.

4a

Verify that the information displayed on the Boot screen is as described by

section 3.3.2.1 on page 3-5

.

4b

Verify that the System Overview screen (see

section 3.3.2.4 on page 3-8

)

appears after the Boot screen.

5

If the information displayed on the Boot screen is not as described by

section

3.3.2.1

, or if the System Overview screen does not appear, contact INFICON.

6

Power down the Sensor Control Unit by setting the power switch to the
0 (OFF) position.