6 sensor screen – INFICON Composer Elite Gas Concentration Monitor User Manual
Page 50
3 - 10
PN
07
4-
56
6-
P1
B
Composer Elite Operating Manual
3.3.2.6 Sensor Screen
On the Overview screen, press Sensor Data to display the Sensor screen (see
). Data displayed on the Sensor screen is for an individual sensor. Data
includes Concentration (Conc), FACTORY ZERO status, Temp 1 and Temp 2.
Indicators (described in
) display when appropriate.
Errors
,
Warnings
or
Errors & Warnings
may be displayed, depending on
whether errors, warnings, or both are present. If no errors or warnings are present,
this area will remain blank.
Zero Reference, Reset Reference, Overview, Parameters, and Advanced Data
are available.
Figure 3-6 Sensor screen
During
, pressing Zero Reference sets a new Zero Reference point.
Zero Reference is only available if the system meets the
AT TEMP
and
STEADY
conditions and if ALLOW USER ZERO is set to YES on the Sensor Parameters
screen.
Reset Reference re-establishes the Factory Zero setting and is only available
when User Zero is displayed. Both buttons appear
grey
when they are
not available.
If Factory Zero is used, FACTORY ZERO is displayed.
If the sensor parameter ALLOW USER ZERO is set to NO, or if the Composer Elite
Sensor Control Unit is not connected to a Composer Elite Acoustic Sensor, (USER
ZERO NOT AVAILABLE) will be displayed next to FACTORY ZERO.
If a user-defined zero is being used, USER ZERO is displayed.