3.3.2.6 Sensor Screen

On the Overview screen, press Sensor Data to display the Sensor screen (see

Figure 3-6

). Data displayed on the Sensor screen is for an individual sensor. Data

includes Concentration (Conc), FACTORY ZERO status, Temp 1 and Temp 2.

Indicators (described in

section 3.3.2.2 on page 3-6

) display when appropriate.

Errors

,

Warnings

or

Errors & Warnings

may be displayed, depending on

whether errors, warnings, or both are present. If no errors or warnings are present,
this area will remain blank.

Zero Reference, Reset Reference, Overview, Parameters, and Advanced Data
are available.

Figure 3-6 Sensor screen

During

Calibration

, pressing Zero Reference sets a new Zero Reference point.

Zero Reference is only available if the system meets the

AT TEMP

and

STEADY

conditions and if ALLOW USER ZERO is set to YES on the Sensor Parameters
screen.

Reset Reference re-establishes the Factory Zero setting and is only available
when User Zero is displayed. Both buttons appear

grey

when they are

not available.

If Factory Zero is used, FACTORY ZERO is displayed.

If the sensor parameter ALLOW USER ZERO is set to NO, or if the Composer Elite
Sensor Control Unit is not connected to a Composer Elite Acoustic Sensor, (USER
ZERO NOT AVAILABLE)
will be displayed next to FACTORY ZERO.

If a user-defined zero is being used, USER ZERO is displayed.