2 sensor 1-5 data set – INFICON Composer Elite Gas Concentration Monitor User Manual
Page 79

5 - 13
PN
07
4-
56
6-
P1
B
Composer Elite Operating Manual
5.4.2.2.2 Sensor 1-5 Data Set
All of the basic and advanced data associated with a single sensor. The data set
can be requested for any of the 5 possible sensors.
Table 5-13 Sensor 1-5 data set
Name
Range
Data Type Note
Cached Data Set
1-5
Byte
Byte
Boards Present
Indicator
0 – 0x1F
Byte
Byte
SS, LS, MH1/PC,
MH2/UZ, AT, CS,
ER, RS, BM, WA
Refer to
Integer
Status bits Refer to
.
SS bit is not relevant in the data set format.
Measurement
Mode
IDLE = 0
READY =1
SEARCH = 2
TRACK = 3
QTRACK = 4
BASELINE = 6
Encoded
Current System Measurement Mode.
Concentration
0.0 – 100.0
mole %
Float
The current concentration value.
Temperature 1
30.0 – 65.0 C
Float
The current temperature value of first
temperature sensor.
Temperature 2
30.0 – 65.0 C
Float
The current temperature value of second
temperature sensor.
Errors
------
Integer
Current error value. Refer to
for a
complete description of the errors.
Warnings
------
Integer
Current warning value See
for a
complete description of the warnings.
Heater 1 Status
0 - 6
Short
0 = Off
1 = Not at temp
2 = At temp
3 = Failed
4 = Cable Disconnected
5 = Near Temp
6 = Over Temp
Heater 2 Status
0 - 6
Short
0 = Off
1 = Not at temp
2 = At temp
3 = Failed
4 = Cable Disconnected
5 = Near Temp
6 = Over Temp
Sample Number
0 – 255
Byte
The current sample number
Frequency
Float
Amplitude
Float
Heater 1 Current
Drawn
Float
Heater 2 Current
Drawn
Float
R Squared
Float