beautypg.com

2 sensor 1-5 data set – INFICON Composer Elite Gas Concentration Monitor User Manual

Page 79

background image

5 - 13

PN

07

4-

56

6-

P1

B

Composer Elite Operating Manual

5.4.2.2.2 Sensor 1-5 Data Set

All of the basic and advanced data associated with a single sensor. The data set
can be requested for any of the 5 possible sensors.

Table 5-13 Sensor 1-5 data set

Name

Range

Data Type Note

Cached Data Set

1-5

Byte

Byte

Boards Present

Indicator

0 – 0x1F

Byte

Byte

SS, LS, MH1/PC,

MH2/UZ, AT, CS,

ER, RS, BM, WA

Refer to

Table 5-18

Integer

Status bits Refer to

Table 5-18

.

SS bit is not relevant in the data set format.

Measurement

Mode

IDLE = 0

READY =1

SEARCH = 2

TRACK = 3

QTRACK = 4

BASELINE = 6

Encoded

Current System Measurement Mode.

Concentration

0.0 – 100.0

mole %

Float

The current concentration value.

Temperature 1

30.0 – 65.0 C

Float

The current temperature value of first
temperature sensor.

Temperature 2

30.0 – 65.0 C

Float

The current temperature value of second
temperature sensor.

Errors

------

Integer

Current error value. Refer to

Table 5-22

for a

complete description of the errors.

Warnings

------

Integer

Current warning value See

Table 5-23

for a

complete description of the warnings.

Heater 1 Status

0 - 6

Short

0 = Off
1 = Not at temp
2 = At temp
3 = Failed
4 = Cable Disconnected
5 = Near Temp
6 = Over Temp

Heater 2 Status

0 - 6

Short

0 = Off
1 = Not at temp
2 = At temp
3 = Failed
4 = Cable Disconnected
5 = Near Temp
6 = Over Temp

Sample Number

0 – 255

Byte

The current sample number

Frequency

Float

Amplitude

Float

Heater 1 Current

Drawn

Float

Heater 2 Current

Drawn

Float

R Squared

Float