Table 5-9 Instance #4 (COS/Cyclic Producing)
#
Name
Access
Type
Comments
1
State
Get
USINT
0 = Nonexistent
1 = Configuring
2 = Waiting for Connections ID
3 = Established
4 = Timeout
2
Instance Type
Get
USINT
0001h (I/O Connection)
3
Transport Class Trigger
Get
BYTE
00h = Client, Cyclic, Class 0, No Ack.
10h = Client, COS, Class 0, No Ack.
02h = Client, Cyclic, Class 2, Ack.
12h = Client, COS, Class 2, Ack.
4
Produced Connection ID
Get
UINT
CAN ID for Transmission
5
Consumed Connection ID
Get
UINT
FFFFh = No Ack.
Other = CAN ID for reception (Ack.)
6
Initial Comm Characteristics
Get
BYTE
0Fh = Producing only over message,
group 1 (No Ack.)
01h = Produces over message, group 1
- Consumes over message, group 2 (Ack.)
7
Produced Connection Size
Get
UINT
Corresponds to Fieldbus Input size
8
Consumed Connection Size
Get
UINT
0000h (consumes 0 bytes on this
connection)
9
Expected Packet Rate
Get/Set(a)
UINT
Timing associated with this connection
12
Watchdog Timeout Action
Get/Set(a)
USINT
0000h (Transition to the timed out state)
13
Produced Connection Path Length
Get
UINT
0007h
14
Produced Connection Path
Get
EPATH
20 04 25 96 00 30 03h
15
Consumed Connection Path
Length
Get
UINT
0000h (No Ack.)
0005h (Ack.)
16
Consumed Connection Path
Get
EPATH
No Value (No Ack.)
20 2B 25 01 00h (Ack.)
17
Production Inhibit Time
Get
UINT
0000h