INFICON Composer Elite Gas Concentration Monitor User Manual
Page 74
5 - 8
PN
07
4-
56
6-
P1
B
Composer Elite Operating Manual
Table 5-7 Instance #2 (Poll or "COS/Cyclic Consuming")
#
Name
Access
Type
Comments
1
State
Get
USINT
0 = Nonexistent
1 = Configuring
2 = Waiting for Connections ID
3 = Established
4 = Timeout
2
Instance Type
Get
USINT
0000h (I/O Connection)
3
Transport Class Trigger
Get
BYTE
82h = Server, Polled, Class 2
80h = Server, COS/Cyclic, Class 0,
No Ack.
82h = Server, COS/Cyclic, Class 2, Ack.
4
Produced Connection ID
Get
UINT
FFFFh = Not Consuming (COS/Cyclic)
Other = CAN ID for transmission
5
Consumed Connection ID
Get
UINT
-
6
Initial Comm Characteristics
Get
BYTE
01h = Polled
- Produces over message, group 1
- Consumes over message, group 2
F1h = COS/Cyclic, No Ack.
- Consumes over message, group 2
01h = COS.Cyclic, Ack.
-Produces over message, group 1
- Consumes over message, group2
7
Produced Connection Size
Get
UINT
0000h = COS/Cyclic
other = Size of Fieldbus Input data (polled)
8
Consumed Connection Size
Get
UINT
Size of Fieldbus Output data
9
Expected Packet Rate
Get/Set(a)
UINT
-
12
Watchdog Timeout Action
Get/Set(a)
USINT
0000h (Transition to the timed out state)
13
Produced Connection Path Length
Get
UINT
0000h (COS/Cyclic)
0007h (Polled)
14
Produced Connection Path
Get
EPATH
No Value (COS/Cyclic)
20 04 25 64 00 30 03h (polled)
15
Consumed Connection Path
Length
Get
UINT
0007h
16
Consumed Connection Path
Get
EPATH
20 04 25 96 00 30 03h
17
Production Inhibit Time
Get
UINT
0000h