INFICON Composer Elite Gas Concentration Monitor User Manual
Page 91
5 - 25
PN
07
4-
56
6-
P1
B
Composer Elite Operating Manual
UZ
Zero Reference (Bit 28)
0|1
Bit
When sensor number is 1-5:
0= Sensor is using Factory Zero.
1= Sensor is using User Zero.
When Sensor Number is 0:
0= All the sensors are using the Factory
Zero.
1= At least one of the sensors is using the
User Zero.
AT
At Temperature (Bit 27)
0|1
Bit
When sensor number is 1-5:
0= Sensor is not at temperature (neither
heater is at temperature).
1= Sensor is at temperature (heater 1 within
+/-0.002ºC and heater 2 within +/-0.005°C).
When Sensor Number is 0:
0= Not all of the online sensors are at
temperature.
1= All the online sensors are at temperature.
CS
Concentration Steady
(Bit 26)
0|1
Bit
When sensor number is 1-5:
0= Sensor Concentration is currently not
steady.
1= Sensor Concentration is currently steady.
When Sensor Number is 0:
0= Not all the online sensors currently have
steady concentration.
1= All the online sensors currently have a
steady concentration.
Reserved (Bit 25)
ER
Errors (Bit 24)
0|1
Bit
When sensor number is 1-5:
0= Sensor currently does not have any
errors.
1= Sensor currently does have errors.
When Sensor Number is 0:
0= None of the sensors currently have
errors.
1= At least one of the sensors currently have
errors.
RS
Reset (Bit 23)
0|1
Bit
0= There has not been a power cycle/reset.
1= There has been a power cycle since the
previous command.
When sent sensor to controller:
bit indicates sensor card reset.
When sent control to external host:
bit indicates controller card reset or a box
power cycle. (bit cleared after returned).
Table 5-18 Status bit (continued)
Status
Bit
Name
Value
Data
Type
Note