3 message router (02h), 4 devicenet object (03h) – INFICON Composer Elite Gas Concentration Monitor User Manual
Page 71
5 - 5
PN
07
4-
56
6-
P1
B
Composer Elite Operating Manual
5.4.1.3 Message Router (02h)
Not used.
5.4.1.4 DeviceNet Object (03h)
DeviceNet objects are shown in
. The supported services are:
Class - Get Attribute Single (Revision, UINT, 0002h)
Instance - Get/Set Attribute Single, Allocate Master/ Slave Connection Set
(4Bh), Release Group 2 Identifier Set (4Ch)
Table 5-3 Instance #1 Attributes (DeviceNet Object)
#
Name
Access
Type
Comments
1
MAC ID
Get/Set(a)
UINT
Currently used MAC ID
2
Baud Rate
Get/Set(a)
UINT
Currently used Baud Rate
0 = 125 kbps
1 = 250 kbps
2 = 500 kbps
3
BOI
Get/Set
UINT
False
4
Bus Off Counter
Get
USINT
00h
5
Allocation Information
Get
Struct.
of:
BYTE,
USINT
(BYTE) Allocation of choice byte
(USINT) MAC ID of master
10
Quick Connect
Get/Set
BOOL
(True) Disable Quick Connect (default)
(False) Enable Quick Connect
100
Disable Auto Baud (b)
Set
BOOL
(True) Disable Auto Baud (default)
(False) Enable Auto Baud
(a) Attribute may not be able to be set if configuration does not allow it.
(b) This setting is stored in non-volatile memory.
- TGF10 Tracer Gas Filler (36 pages)
- Sensistor ILS500 F Leak Detection Filler (90 pages)
- T-Guard Leak Detection Sensor (85 pages)
- T-Guard Leak Detection Sensor Interface description (40 pages)
- Sensistor ISH2000P Hydrogen Leak Detector, Panel Model (51 pages)
- Sensistor ISH2000 HySpeed Hydrogen Leak Detector (54 pages)
- LDS3000 Modular Leak Detector (52 pages)
- LDS3000 Modular Leak Detector Interface description (56 pages)
- BM1000 Bus module (14 pages)
- I/O1000 I/O module (18 pages)
- CU1000 Control unit (24 pages)
- Helium Leak Detector Modul1000 (130 pages)
- Helium Leak Detector Modul1000 Interface description (40 pages)
- UL5000 Dry Helium Leak Detector (108 pages)
- UL5000 Dry Helium Leak Detector Interface description (14 pages)
- UL1000 Fab Dry Helium Leak Detector (119 pages)
- HLD6000 Refrigerant Leak Detector (76 pages)
- HLD6000 Refrigerant Leak Detector Interface Description (40 pages)
- IO1000 I/O module (18 pages)
- Ecotec E3000 Multigas-Sniffer-Leak Detector (92 pages)
- Ecotec E3000 Multigas-Sniffer-Leak Detector Interface description (36 pages)
- Sensistor XRS9012 Hydrogen Leak Detector User Manual (28 pages)
- Sensistor XRS9012 Hydrogen Leak Detector Maintenance manual (14 pages)
- Extrima Ex-certified Hydrogen Leak Detector (62 pages)
- Sensistor ILS500 Leak Detection System (107 pages)
- Sensistor ISH2000 Hydrogen Leak Detector (58 pages)
- Sensistor ISH2000 Hydrogen Leak Detector (108 pages)
- Sensistor Sentrac Hydrogen Leak Detector (86 pages)
- Protec P3000(XL) Helium Leak Detector (132 pages)
- Pilot Plus Vacuum Gauge (2 pages)
- CO Check Carbon Monoxide Meter (2 pages)
- GAS-Mate Combustible Gas Leak Detector (12 pages)
- Whisper Ultrasonic Leak Detector (8 pages)
- Vortex AC Refrigerant Recovery Machine 115V (20 pages)
- Vortex AC Refrigerant Recovery Machine 230V (16 pages)
- Wey-TEK Refrigerant Charging Scale & Optional Charging Module (2 pages)
- Wey-TEK Refrigerant Charging Scale & Optional Charging Module (44 pages)
- D-TEK CO2 Refrigerant Leak Detector (12 pages)
- TEK-Mate Refrigerant Leak Detector (12 pages)
- Compass Refrigerant Leak Detector (12 pages)
- D-TEK Select Refrigerant Leak Detector (12 pages)
- Explorer Portable Gas Chromatograph (369 pages)
- MicroFID II Portable Flame Ionization Detector (89 pages)
- DataFID Portable Flame Ionization Detector for Landfill Emissions Monitoring (91 pages)
- Hydrostik Hydrogen Fuel Cylinder Installation (7 pages)