4 sensors 1,2,3 error/warning data set – INFICON Composer Elite Gas Concentration Monitor User Manual
Page 84
5 - 18
PN
07
4-
56
6-
P1
B
Composer Elite Operating Manual
5.4.2.2.4 Sensors 1,2,3 Error/Warning Data Set
This data set returns a compound set of current data values for three of the five
sensor cards.
Table 5-15 Sensors 1,2,3 error/warning data set
Name
Range
Data Type Note
Cached Data Set
7
Byte
Numeric value of the data set cached.
Boards Present
Indicator
0 – 0x1F
Byte
Bitwise: Bits 0 – 4 set indicate sensor board
1 – 5 present, respectively.
Sensor 1
SS, LS, MH1/PC,
MH2/UZ, AT, CS,
ER, RS, BM, WA
Refer to
Integer
Status bits Refer to
SS bit is not relevant in the data set format.
Sensor 1
Errors
------
Integer
Current error value Refer to
for a
complete description of the errors.
Sensor 1
Warnings
------
Integer
Current warning value Refer to
a complete description of the warnings.
Sensor 1
Sample Number
0 – 255
Byte
The current sample number.
Sensor 1
Frequency
Float
Sensor 1
Amplitude
Float
Sensor 1
Heater 1 Current
Drawn
Float
Sensor 1
Heater 2 Current
Drawn
Float
Sensor 2
SS, LS, MH1/PC,
MH2/UZ, AT, CS,
ER, RS, BM, WA
Refer to
Integer
Status bits Refer to
.
SS bit is not relevant in the data set format.
Sensor 2
Errors
------
Integer
Current error value. Refer to
for a
complete description of the errors.
Sensor 2
Warnings
------
Integer
Current warning value Refer to
a complete description of the warnings.
Sensor 2
Sample Number
0 – 255
Byte
The current sample number.
Sensor 2
Frequency
Float
Sensor 2
Amplitude
Float
Sensor 2
Heater 1 Current
Drawn
Float
Sensor 2
Heater 2 Current
Drawn
Float