beautypg.com

Chapter 2 installation, 1 composer elite installation guidelines, 1 pressure, temperature, and tubing – INFICON Composer Elite Gas Concentration Monitor User Manual

Page 29: 1 maximum pressure, Chapter 2

background image

2 - 1

PN

07

4-

56

6-

P1

B

Composer Elite Operating Manual

Chapter 2

Installation

2.1 Composer Elite Installation Guidelines

The requirements and recommendations in

section 2.1.1

,

section 2.1.2

, and

section 2.1.3

must be reviewed and understood before Composer Elite is installed

as described by

section 2.2, Composer Elite Installation, on page 2-8

.

2.1.1 Pressure, Temperature, and Tubing

2.1.1.1 Maximum Pressure

Reliable measurement of concentration to absolute pressures as low as 26.66 kPa
(200 Torr) requires relatively thin and flexible diaphragms to be used so that
sufficient acoustic energy may be transmitted.

There is a possibility that the diaphragms may destructively rupture if an absolute
pressure greater than 826 kPa (6200 Torr) is applied. A secondary containment
chamber is built into the Acoustic Sensor to prevent process gases from escaping
the Acoustic Sensor in the event of a diaphragm rupture.

CAUTION

Do not connect the Acoustic Sensor to process tubing
that can exceed an absolute pressure of
826 kPa (6200 Torr). Exposing the Acoustic Sensor to
excess pressure may cause catastrophic diaphragm
failure or may stretch the diaphragms, requiring on-site
reference zeroing (see

section 4.1.2 on page 4-1

).


A certified rupture disk should be installed upstream of
the Acoustic Sensor to limit any possible over-pressure
to less than 826 kPa (6200 Torr).