2 devicenet data sets, 1 initialization, 2 data – INFICON Composer Elite Gas Concentration Monitor User Manual
Page 78: 1 data sets
5 - 12
PN
07
4-
56
6-
P1
B
Composer Elite Operating Manual
Table 5-11 Instance Attributes (01h)
5.4.2 DeviceNet Data Sets
5.4.2.1 Initialization
After a power cycle, Composer Elite will default to expecting and responding to
INFICON messages. To instruct Composer Elite to cache data in CE OUT instead,
in the host computer set the I/O Data Set attribute to a non-zero value. The value
will specify to Composer Elite which data set to cache. This value can be changed.
When Composer Elite is caching data to CE OUT, CE IN is not used.
NOTE: Switching data sets, or between INFICON messages and a data set, may
take a few moments to complete.
5.4.2.2 Data
5.4.2.2.1 Data Sets
Data sets are preceded by a Count, an Update In Progress flag and a Cached Data
Set value.
The Cached Data Set value has been repeated in the following byte
descriptions for improved clarity.
#
Name
Access
Type
Comments
100
I/O Data Format
Get/Set
UINT
0 = INFICON Messaging
1 = Sensor 1 Data Set
2 = Sensor 2 Data Set
3 = Sensor 3 Data Set
4 = Sensor 4 Data Set
5 = Sensor 5 Data Set
6 = All 5 Sensor Card Data
Set
7 = Sensors 1,2,3 Err/Wrn
Data Set
8 = Sensors 4,5 Err/Wrn
Data Set
9 = All 5 Sensor Card
Concentration
Table 5-12 Cached data sets values byte descriptions
Value
Size (bits)
Field
Type
Definition
Count
-----------
16
Short
Value that when monitored indicates when
data values may have changed.
Update in
Progress Flag
0-1
8
Byte
TRUE if the data values are being updated.
FALSE otherwise. Used to determine if a
partial value could be present.
Cache Data
Set
1-9
9
Byte
Indicates how to interpret the following bytes
of data.
Composer
Elite Data
Values
-----------
(N-3) * 8
bits
Byte
Data values. See Data Set tables.