Multi-film density parameter, Multi-film z-factor parameter, Multi-film end thickness parameter – INFICON STM-100/MF Thin Film Deposition Monitor User Manual

Page 69: Multi-film set point relay thickness parameter, Multi-film set point timer relay parameter, Multi-film tooling factor parameter, Computer interfacing page 5 - 14

background image

_________________________________________________________________________

Command:

j

Argument:

[1 thru 9] [=0.500 .. 99.99] l [?]

Description:

Sets the density parameter for the film specified in the first argument. If a "?" is
sent as a parameter, the current density value is returned. Example: command
j6=0.75 sets the density for film 6 to 0.750 gm/cc. The units are gm/cc.

_________________________________________________________________________

Command:

k

Argument::

[1 thru 9] [=0.100 .. 99.99] l [?]

Description:

Sets the material Z-Factor for the film specified in the first argument. If a "?" is the
parameter, the current Z-Factor is returned. Example: command k5=0.5 sets the Z-
Factor for film 5 to 0.5. Dimensionless number.

_________________________________________________________________________

Command:

l

Argument:

[1 thru 9] [=0 .. 9999000] l [?]

Description:

Sets the end thickness for the film specified in the first argument. While any
resolution number can be programmed (up to 7 digits), the number that is accepted
by the STM-100 / MF will be within the display range. This is limited to a total of 4
digits. The units are Angstroms.

_________________________________________________________________________

Command:

m

Argument:

[1 thru 9] [=0 .. 9999000] l [?]

Description:

Sets the Set Point Relay thickness limit for the film specified in the first argument.
See the description for the G command.

_________________________________________________________________________

Command:

n

Argument:

[1 thru 9] [=00:00 .. 99:59] l [?]

Description:

Set the setpoint timer relay for the film specified in the first argument. The format is
"Minutes: Seconds". A "?" will return the current value.

_________________________________________________________________________

Command:

o

Argument:

[1 thru 9] [=10.0 .. 399] l [?]

Description:

Set the tooling factor parameter for the film specified in the first argument. A "?" will
return the current value. The units are in percent.

Multi-Film Density
Parameter

Multi-Film Z-Factor
Parameter

Multi-Film End
Thickness Parameter

Multi-Film Set Point
relay Thickness
Parameter

Multi-Film Set Point
Timer Relay Parameter

Multi-Film Tooling
Factor Parameter

COMPUTER INTERFACING

Page 5 - 14